ON SITE

Instrument Calibrated/
Measurement Parameter
Range Calibration and Measurement
Capability Expressed as an
Uncertainty (±)*
Remarks
PRESSURE
Pressure measuring instruments 15 psi to 3000 psi
3000 psi to 5000 psi
1 psi
2 psi
Calibrate using dead weight tester as standards based on DKD-R 6-1 sequence A, C
FORCE
Push pull gauge / digital force gauge / tension gauge 0 kgf to 0.5 kgf
0.5 kgf to 1 kgf
1 kgf to3 kgf
3 kgf to 50 kgf
50 kgf to 100 kgf
0.001 kgf
0.002 kgf
0.005 kgf
0.01 kgf
0.1 kgf
Calibrate using standard weight and poise weights. Calibrations may be given in other units by conversion from Sl units
MASS
Standard Weight 1 g, 2g, 5g
10 g, 20 g, 50 g
100 g
200 g
500 g
1 kg
2 kg
5 kg
10 kg
20 kg
0.1 mg
0.1 mg
0.2 mg
0.2 mg
0.001 g
0.01 g
0.01 g
0.02 g
0.1 g
0.2 g
Calibrate using reference standard weight by comparison method according to ABBA weighing scheme
TEMPERATURE
Temperature indicating
instrument
Type K
Type J
Pt 100


-30 °C to 1200 °C
-30 °C to 1200 °C
-30 °C to 800 °C


0.4 °C
0.4 °C
0.2 °C
Calibrate by electrical simulation using temperature calibrator

Download Scope Of Accreditation in pdf

LABORATORY

Instrument Calibrated/
Measurement Parameter
Range Calibration and Measurement
Capability Expressed as an
Uncertainty (±)*
Remarks
PRESSURE
Pressure measuring instruments 15 psi to 3000 psi
3000 psi to 5000 psi
1 psi
2 psi
Calibrate using dead weight tester as standards based on DKD-R 6-1 sequence A, C
FORCE
Push pull gauge / digital force gauge / tension gauge 0 kgf to 0.5 kgf
0.5 kgf to 1 kgf
1 kgf to3 kgf
3 kgf to 50 kgf
50 kgf to 100 kgf
0.001 kgf
0.002 kgf
0.005 kgf
0.01 kgf
0.1 kgf
Calibrate using standard weight and poise weights. Calibrations may be given in other units by conversion from Sl units
MASS
Standard Weight 1 g, 2g, 5g
10 g, 20 g, 50 g
100 g
200 g
500 g
1 kg
2 kg
5 kg
10 kg
20 kg
0.1 mg
0.1 mg
0.2 mg
0.2 mg
0.001 g
0.01 g
0.01 g
0.02 g
0.1 g
0.2 g
Calibrate using reference standard weight by comparison method according to ABBA weighing scheme
TEMPERATURE
Temperature indicating
instrument
Type K
Type J
Pt 100


-30 °C to 1200 °C
-30 °C to 1200 °C
-30 °C to 800 °C


0.4 °C
0.4 °C
0.2 °C
Calibrate by electrical simulation using temperature calibrator

Download Scope Of Accreditation in pdf